NOTE:
Inspect the evaporator temperature sensor when it is installed to the A/C unit.
1. Set the fan speed at MAX HI.
2. Set the temperature control at MAX COLD.
3. Set the RECIRCULATE mode.
4. Turn the A/C switch off.
5. Close all doors and windows.
6. Wait for 5 min
.
7. Disconnect the evaporator temperature sensor connector.
8. Measure the temperature at the blower inlet.
9. Measure the resistance between the evaporator temperature sensor terminals.

If the resistance is not as shown in the graph, replace the evaporator temperature sensor.
Evaporator Temperature Sensor Inspection [Full Auto Air Conditioner]
Evaporator Temperature Sensor [Full Auto Air Conditioner]Starting the Engine
WARNING
Radio waves from the key may affect medical devices such as pacemakers: Before
using the key near people who use medical devices, ask the medical device manufacturer
or your physician if radio waves from the key will affect the device.
NOTE
• The key must be carried because the
ke ...
Instrumentation/Driver Info. Personalization
A personalization function has been adopted which enables the instrument
cluster settings to be changed.
The Mazda Modular Diagnostic System (M-MDS) is required to perform the setting.
Refer to the Workshop Manual for the detailed setting procedure.
Instrument cluster
...
Hazard Warning Switch Inspection
1. Disconnect the negative battery cable..
2. Remove the decoration panel..
3. Verify that the continuity between the hazard warning switch terminals is
as indicated in the table.
If the continuity is as indicated in the table, go to the next step.
If not as indicated i ...